Operations Manual
HP2000
With 9100 Controller
Self-Heated Oxygen
Measurement System
Please read, understand, and follow these instructions before operating this equipment.
Super Systems, Inc. is not responsible for damages incurred due to a failure to comply with these instructions. If at
any time there are questions regarding the proper use of this analyzer, please contact us at (800) 666-4330 for
assistance.
7205 Edington Drive
Cincinnati, OH 45249
513-772-0060 / 800-666-4330
Fax: 513-772-9466
www.supersystems.com
Page view 0
1 2 3 4 5 6 7 8 9 10

Summary of Contents

Page 1 - With 9100 Controller

Operations ManualHP2000With 9100 ControllerSelf-Heated OxygenMeasurement SystemPlease read, understand, and follow these instructions before operating

Page 2

Spare Parts List• CeraGold Probe with “S” T/C Part Number 10138• Sample Pump Part Number 37134• 110VAC Heater Part Number 32074• Large Bowl Filer

Page 3

Electrical ConnectionsRemove the screw cover on the electrical enclosure box on the side of the main box. Route thegreen “S” Type thermocouple wire a

Page 4

Once the probe has been inserted into the heater, make the necessary electrical connections at thehead of the probe. The green thermocouple wire from

Page 5

Entering Values Through the KeypadOnly numeric values may be entered into the HP2000 display. The up arrow will cycle throughthe numbers starting with

Page 6

Screen 1: Oxygen (in %), Probe mV and Probe TemperatureThis screen is the default screen for the HP2000. This screen lists the percentage of oxygen o

Page 7

This screen will display the temperature set point for the HP2000, and it will allow the user tomodify the temperature set point for the HP2000. The H

Page 8

This screen will display the temperature mode and it will also allow the user to change thetemperature mode. The temperature mode can be either Fahre

Page 9

The Output 1 Exponent provides greater detail for the output. The valid range of values is zeroto ten. To change the value, press the up arrow key o

Page 10

The valid range of values for output 2 full scale is zero to nine thousand nine hundred ninetynine. To change the value, press the up arrow key once